Table 6–4: Failure symptoms and possible causes (Cont.)
Symptom
Possible cause(s)
Flat LCD panel display blank
H
External monitor plugged in but not powered on.
H
External monitor plugged in (look on external monitor for
display).
H
Connect an external monitor cable to the SCOPE ONLY
VGA OUT rear panel connector. If monitor has a display,
replace the LCD panel assembly.
H
Connect an external monitor cable to the SVGA OUT rear
panel connector, if monitor has a display the BIOS may
need to be setup.
H
If the BIOS is OK then replace the Power PC circuit board
assembly.
H
Connect an external monitor cable to the SVGA OUT rear
panel connector. If the monitor has no display, replace the
NLX circuit board assembly.
H
If the monitor has no display when connected to the SVGA
OUT rear panel connector, replace the NLX circuit board
assembly.
H
Defective cable from front panel board to riser board.
H
Defective cable from Power PC board to front panel board.
Front panel LEDs do not light.
H
Replace the front panel board.
Hard disk drive related
symptoms
H
Defective hard disk drive.
H
Incorrect hard disk type selected in the BIOS setup.
H
Replaceable hard disk drive not installed.
H
Power supply failure.
H
Corrupted controller BIOS module firmware, reinstall
firmware.
H
Controller BIOS setup problem. See BIOS Settings.
H
Replaceable hard disk drive or optionally field installed fixed
hard disk drive not configured as bootable (slave) master
hard disk drive.
H
Faulty benchtop controller.
CD-ROM related symptoms
H
Defective CD-ROM.
H
Defective CD-ROM drive cable.
H
Defective CD-ROM board.
H
Incorrect CD-ROM configuration in the BIOS setup.
Содержание CSA8000 Series
Страница 4: ......
Страница 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 23: ......
Страница 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Страница 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 64: ......
Страница 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 93: ......
Страница 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 104: ......
Страница 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 219: ......
Страница 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 308: ......
Страница 313: ......
Страница 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 316: ......