Installation
CSA8000 & TDS8000 Instruments and Sampling Modules
2–9
2. If a keyboard is not connected to the instrument, connect it.
3. Insert your emergency startup disk in the instrument floppy drive.
4. Power on the instrument. Make sure the principle power switch at the back
of the instrument is set on. Push the On/Standby switch on the front panel.
5. The instrument should now boot from the emergency startup disk and
present you with a selection menu. Use the arrow and enter keys to choose
Start the computer without CDROM support
.
6. At the a:\ prompt, type
c:
. Then type
cd windows\options\cabs
.
7. Type
setup
. Follow the instructions on screen.
NOTE
. At the a:\ prompt, you can type
help
at anytime to read a file that may be
helpful in getting the Windows 98 operating system up and running.
System Hard Drive Rebuild.
If you cannot reboot from the instrument hard drive,
you must rebuild the instrument hard drive. This process will return the hard disk
to the its original condition present when the instrument shipped.
Since data and programs you may have installed are lost with this rebuild, you
should do the Windows 98 Reinstall Only procedure instead if you can. (See
Windows 98 Reinstall Only on page 2–8.)
If you must rebuild the system hard drive, use the System Rebuild CD that
shipped with the instrument. Follow the instructions in the booklet found in the
CD case.
In case of instrument problems, you may wish to run the system diagnostics. If
so, see the procedure Perform the Diagnostics, on page 4–3.
If the instrument is turned off before the operating system boots, or if you’ve
installed a third-party product with a driver incompatible with instrument start
up, Windows 98 will open in Safe mode. The touchscreen will be inoperable;
therefore, you must install the standard-accessory mouse and keyboard to operate
the instrument.
When you have finished investigating and removed any barrier to Windows
start-up, you can reboot. If the instrument no longer boots to Safe mode, you can
remove the keyboard and mouse if desired.
System Diagnostics
Windows Safe Mode
Содержание CSA8000 Series
Страница 4: ......
Страница 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 23: ......
Страница 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Страница 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 64: ......
Страница 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 93: ......
Страница 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 104: ......
Страница 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 219: ......
Страница 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 308: ......
Страница 313: ......
Страница 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 316: ......