11. Calculate the DUT response as the difference between the 80E01 Measured
level column and the Reference RF level column in Table 4–6 as:
(80E01 Measured level) – (Reference RF level)
Record the calculation results in the DUT Response column Table 4–6 on
page 4–71.
12. Calculate all frequencies listed in the Frequency column, Table 4–6 on
page 4–71.
13. Find the maximum and minimum power points between the 50 MHz and
50 GHz points in the DUT Response column, Table 4–6 on page 4–71.
14. The maximum power minus the power at 50 MHz should be greater than
+2.3 dB. The minimum power minus the power at 50 MHz should be greater
than –2.3 dB. The specification is
"
3.0 dB referenced to the 50 MHz.
End of Procedure
Содержание CSA8000 Series
Страница 4: ......
Страница 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 23: ......
Страница 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Страница 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 64: ......
Страница 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 93: ......
Страница 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 104: ......
Страница 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 219: ......
Страница 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 308: ......
Страница 313: ......
Страница 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 316: ......