5. Set the synthesized signal generator frequency to 2.5 GHz.
6. Set the Synthesized output to 1600 mV peak-to-peak (800 mV displayed)
7. Set up the CSA8000/TDS8000 as follows:
a. Select Default Setup from the Utilities menu on the menu bar at top of
display.
b. Select C1 from the Waveform Selector menu on the controls bar at the
bottom of the display (left corner).
c.
In the Trig Setup dialog box:
H
select External Prescaler.
H
select Metastablility Reject.
H
select Auto Mode.
d. In the Vert Setup dialog box:
H
set Scale to 100 mV/division.
H
set Offset to 0.
e.
In the Horz Setup dialog box:
H
set Scale to 200 ps/division.
H
set Horizontal Reference to 0%.
H
set Position to 19 ns.
H
set All Timebases Mode to Short Term Jitter.
Follow this procedure to make the external prescale check:
1. In the Meas Setup dialog box:
a. make sure the channel under test is selected in the Source menu.
b. select Meas 1, and then choose Select Meas > Amplitude > Amplitude.
c.
select Meas 2, and then choose Select Meas > Eye Pattern/Optical >
RMS Jitter. Dismiss the database notification that pops up.
d. check the Use Wfm Database box in the Source tab.
e.
check Display Database in the Wfm Database dialog box.
2. Adjust the synthesizer signal generator level until the amplitude measure-
ment readout on the instrument screen indicates the amplitude is 800 mV.
Test:
Содержание CSA8000 Series
Страница 4: ......
Страница 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 23: ......
Страница 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Страница 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 64: ......
Страница 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 93: ......
Страница 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 104: ......
Страница 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 219: ......
Страница 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 308: ......
Страница 313: ......
Страница 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 316: ......