Adjustment Procedures
CSA8000 & TDS8000 Instruments and Sampling Modules
5–3
Main Instrument Adjustments
Once the instrumentation is connected and GPIB configured, allow the instru-
mentation to warm up for at least 20 minutes before starting the DC Calibrator
Adjust procedures.
Digital
multimeter
–
+
DC CALIBRATION OUTPUT
CSA8000/TDS8000
50
W
Terminator
SMA “T”
Controller
GPIB cable
Figure 5–1: Adjustment setup using the DMM
Connect the instrument to a DMM as shown in Figure 5–1.
1. Run the compensation routine, described in Brief Procedures section on
page 4–5.
2. Enter the following GPIB command to verify communication between the
controller and DUT.
*IDN?
Should respond with Tektronix and Firmware Version.
3. Enter the following GPIB command to turn the instrument’s cal constant
protection off:
SYST:PROT OFF
4. Enter the following GPIB command to set the instrument’s DC calibrator
offset cal constant to 0:
CALCOMP:DOUBLE “DcCalOffsetAdj”,0.0
5. Wait 8 seconds then enter the following GPIB command to set the
instrument’s DC calibrator Lsb cal constant
to 1.0:
CALCOMP:DOUBLE “DcCalLsbAdj”,1.0
6. Wait 8 seconds then enter the following GPIB command to set the
instrument’s DC calibrator output to 0 Volts:
DC Calibrator Adjust
Содержание CSA8000 Series
Страница 4: ......
Страница 18: ...Service Safety Summary xiv CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 22: ...Preface xviii CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 23: ......
Страница 59: ...80C00 Modules Specifications 1 36 CSA8000 TDS8000 Instruments and Sampling Modules back ...
Страница 63: ...80A01 Module Specifications 1 40 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 64: ......
Страница 74: ...Installation 2 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 93: ......
Страница 103: ...Theory of Operation 3 10 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 104: ......
Страница 106: ...Performance Verification Procedures 4 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 122: ...Brief Procedures 4 18 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 126: ...Performance Tests 4 22 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 218: ...80A01 Trigger Preamplifier Module 4 114 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 219: ......
Страница 231: ...Maintenance 6 6 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 291: ...Removal and Installation Procedures 6 66 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 307: ...Repackaging Instructions 6 82 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 308: ......
Страница 313: ......
Страница 315: ...Electrical Parts List 8 2 CSA8000 TDS8000 Instruments and Sampling Modules ...
Страница 316: ......