6. Adjust Menu
Adjust Menu
other yes
Plate: 1
- + other yes
Primetime:10 sec
- + other yes
Rinsetime: 10sec
- + other yes
AutoRinse: 1 h
- + other yes
Kit Select: Off
- + other yes
Date: ddmmmyyyy
d m other y
Time: hh:mm xx
h m other form
Sel.Manifold: xx
- + other yes
Confirm selection of Adjust menu
Plate Type definition menu
Use
-
and
+
keys to select required plate type
Plate Type definition procedure
Priming Time setting
Use
-
and
+
keys to select required priming time
(5 - 99 sec.)
Rinse Time setting
Use
-
and
+
keys to select required rinsing time
(5 - 99 sec.)
Automatic Rinsing Time setting
Use
-
and
+
keys to select required time between
Automatic Rinsing procedures (1 - 24 hrs.)
This option is not available with
this instrument
Date setting (option)
Use
d
to change day setting
Use
m
to change month setting
Use
y
to change year setting
Time setting (option)
Use
h
to change hour setting
Use
m
to change minute setting
Use
Form
to change from 12 to 24 hr format
Manifold setting (optional)
Use
-
and
+
keys to select fitted manifold. (8, 12, 16)
If manifold detect option or fill verfication is enabled,
this setting can not be used
Exit Adjust menu
Adj. Liq Sens.:
other yes
Exit Adjust
other yes
Run XX yyyyyyyy
- + other yes
Adjust Liquid Sensor (optional)
Liq. Sens. 1
- + exit yes
Empty 1 XX 1
exit yes
Empty 1 XX 0
exit yes
XX = displayed value
0 = values below limit
Press
exit
to continue the adjustment of other
sensors.
XX = displayed sensor value (does not directly correlate to
the volume of liquid in the bottles).
1 = value above limit
Press
yes
to store the currently displayed value as the limit.
Using the - and + keys select the liquid sensor.
1 = Green (wash solution)
3 = Blue
2 = Gray (rinse solution)
4 = Purple
W = Red (waste)
Press
yes
to continue the adjustment procedure.
6-2
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
May 2003