3. Installation Procedure
3.4 Installation
Procedure
Caution
Before the instrument is installed and switched on,
it should be left to stand for at least three hours, so there is
no possibility of condensation causing a short circuit.
3.4.1
Installing the Instrument
•
Unpack the instrument and check the contents.
See 3.2 Unpacking Procedure
.
•
Connect the solution tubes to the inlet connectors.
See below 3.4.2 Connecting the Solution Tubes
.
•
Setup possible instrument options.
See page 3 – 4 and 3 – 5.
•
Connect instrument to the power supply.
See 3.4.6 Power Connections
.
•
Define the plate parameters in the Adjust menu.
See chapter 6.
•
Program the wash procedure.
See chapter 5
.
3.4.2
Connecting the Solution Tubes
The solution tubes are connected onto the inlet connectors on the back panel of
the instrument. The instrument is supplied with four meters of tubing that can be
cut to the required length (maximum 1 meter).
Fig. 3-2 Connecting the Solution Tubes
Inlets for channels 1 - 4
Waste Tube
red
Aerosol Tube
(optional) yellow
May 2003
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
3-3