7. Maintenance and Cleaning
7.1.3
Exchanging the Plate Support
The instrument is delivered with the plate support already installed. If the plate
support is removed, for example: for cleaning, it can be replaced using the
following procedure:
Lift the manifold arm
Slightly tilt the plate support towards the right.
Carefully slide the plate support with the waste sink at the rear over the guide
bar.
The plate support must be slightly tilted so that it goes over the pin that activates
the automatic microplate centering mechanism.
To reinstall the plate support follow the procedure listed below:
Push the plate support so that it is fully inserted into the instrument.
Lower the plate support.
Lower the manifold arm.
7-6
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
May 2003