3. Installation Procedure
3.4.4
Connecting the Barcode Reader (Optional)
Before the barcode reader can be connected to the instrument, the supplied
barcode chips must be added to the main board. For more details contact your
service engineer.
Fig. 3-4 Connecting the Barcode Reader
Hand Held Barcode
Reader
Liquid Sensor
3.4.5
Connecting the Printer (Optional)
Before the printer can be connected to the instrument, the supplied chip must be
added to the main board. For more details contact your service engineer.
Connect the printer to the instrument using the supplied cable.
Connect the power supply cable to the printer.
Fig. 3-5 Connecting the Printer
PAPER
FEED
BATT. LOW
POWER
POWER
OFF 1 2
ON
Power Cable
ON /OFF
Switch
Connection Cable
3.4.6 Power
Connections
Caution
Before the instrument is installed and switched on,
it should be left to stand for at least three hours, so there is
no possibility of condensation causing a short circuit.
Ensure that the on / off switch in the back panel of the instrument is in the off
position.
Insert the mains power cable into the mains power socket on the rear of the
instrument.
May 2003
Instructions for Use for Columbus Washer No. I 109 006 Rev No. 2.2
3-5