4. Operating Instructions
4.5 Error
Messages
4.5.1 Standard
Instrument
Program not defined
If the selected program has
not
been defined, the following message is displayed:
Program not def
exit
Press
exit
to remove the message and return to the standby mode
Run XX yyyyyyyy
- + other yes
Select a defined program that is to be performed.
Wrong Manifold
If the selected program requires a different type of manifold than the manifold that
is fitted to the instrument, the following message is displayed:
WRONG MANIFOLD
exit
Press
exit
and the instrument returns to the standby mode.
Either exchange the manifold for the required type of manifold or re-program the
procedure for the type of manifold that is fitted.
If the correct type of manifold is fitted, check the setting for the fitted manifold in
the Adjust menu.
If either the Manifold Detection or Fill Verification option is enabled ensure that
the ribbon cable is in the correct connector, see diagram in chapter 2.
This message may also be displayed, if the Fill Verification or Manifold Detection
options are enabled but the required parts are not fitted to the instrument. In this
case please call your local representative.
Transport Error
If the instrument cannot move the plate support, the following message is
displayed:
Init-Err. CONT?
retry yes
Press
yes
(exit)
to remove the error message and the instrument returns to the
standby mode.
Check that the plate support is correctly inserted and that the plate support
system is clear.