SYMEO Local Positioning Radar System
Product Family LPR
®
-1DHP-200 – Product Documentation
Quick Setup
Copyright © Symeo GmbH 2018
DOC.EDO.000341.WORK.EN_LPR-1DHP-200_Product-Documentation.docx
Page 28 of 71
Device -> Settings -> Measurement details
o
Target search mode = “First” if the reflector is the first target in range, “Strongest” if
the reflector is the strongest target in range.
The setting “Target search mode” highly impacts the behavior of the radar,
especially in multi target environments (see Figure 8.16
)
. “First target” may
lead to distance measurements to unintended targets in the vicinity of the
radar (e.g. a person passing the radar beam). “Strongest target” may lead
to distance measurements to unintended targets present in the
background of your intended target (e.g. a wall behind a corner reflector).
Perform therefore always a test as described in chapter 6.4 to verify this
setting.
Review and Calibrate your measurement as stated in chapter 6.4.
6.3 Quick Setup for Secondary Radar Mode
The following settings must be set in the WebUI of the master and slave sensor for operation
in secondary radar mode.
Only the Master unit outputs range data.
Master
Device -> Settings -> Measurement basics
o
Station mode = Master
o
Bandwidth mode = Choose a bandwidth that fits your required range and accuracy
o
Channel block = Use recommended selection
o
Sync channel = different than any other LPR
®
-1DHP-200 sensor in range (except the
related Slave); for neighboring measurement paths use only every fourth sync
channel.
Slave
Device -> Settings -> Measurement basics
o
Station mode = Slave
o
Bandwidth mode = same as Master
o
Channel block = same as Master
o
Sync channel = same as Master
Review and Calibrate your measurement as stated in chapter 6.4.
6.4 Review and Calibration of the Measurement Setup
Diagnostics -> Range measurement statistics -> Live range measurement