49
SVANTEK 977W User Manual
There are two ranges available:
High
and
Low
. The detailed description of the
measurement ranges parameters is given in
App. C. After pressing the
<ENTER>
push-
button the change is confirmed and the
window closes. Press the
<ESC>
push-
button to return to the
Measurement
list
ignoring any changes made in the sub-list.
►
The above screens were made with
calibration factor equal to zero. The
calibration factor is always added to the
upper range level
– see example.
=>
Right-side screens present the
Range
for
VLM mode and zero calibration factor.
►
Note: The calibration factor is always added to the range limits. For
example, if calibration factor is equal to 0.5 dB, the range will be
changed automatically (lower and upper limits will be increased by
0.5).
5.8
Setting the RPM measurements
– RPM
To perform the RPM measurement the RPM probe should be connected to the
I/O
socket. The instrument automatically detects such connection and switches
the
I/O Mode
to
Digital In
(
path: <Menu> / Instrument / Multifunction I/O /
I/O Mode
).
While the tacho is connected the
I/O Mode
position is inactive.
The
RPM Measurement
position enables
the user to programme the measurements
of rotational speed, measured by tacho
probe. The RPM function usually is an
optional function and should be activated
during first attempt to switch it on.
<ENT>