78
SVANTEK 977W User Manual
9. SETTING THE AUXILIARY PARAMETRS
– Auxiliary Setup
The
Auxiliary Setup
list provides the user
with additional features that allow, for
instance, customize the interface device to
a specific user and are not directly related
to the hardware components of the
instrument. To open the
Auxiliary Setup
list, the user should press the
<Menu>
push-button, select the
Auxiliary Setup
text and press
<ENTER>
.
<ENT>
In the
Auxiliary Setup
list, the following items are available:
Language
enables the user to select the language of the user interface.
Factory Settings
enables the user to return to the default, factory settings.
Reference Levels
enables the user to select the reference level for the Vibration measurements and it
informs the user about the reference level in the Sound measurements.
Vibration Units
enables the user to select the Vibration units in which the results of the measurements
are to be viewed. This position appears only in Vibration modes.
Warnings
enables the user to switch the warnings on or off that can be displayed during the normal
operation of the instrument.
9.1. Setting language of the user interface
– Language
The
Language
sub-list enables the user to
select the language of the user interface.
If after turn on the instrument an unknown
language interface appears on the display
the user can reset the instrument by means
of the four <
Shift/Enter/Alt/Start>
push-
buttons pressed together. After this, the
instrument will come back to the default
setup with the English interface.
<ENT>
9.2. Restoring the factory settings
– Factory Settings
The
Factory Settings
sub-list enables the
user to restore the default settings of the
instrument.
Select
Yes
in the
Factory Settings
window
and press
<ENTER>
. After restoration
process the instrument will inform the user
that “
Settings restored
”.
The factory setup can be installed also by
means of the four
<Shift/Enter/Alt/Start>
push-buttons pressed together.
<ENT>