87
SVANTEK 977W User Manual
After the program dump the Bootstrap and
Program files names are presented on
subsequent screens.
<Key>
The user may find the files with the memory dump in the selected directory
among other files.
Note:
The PROGRAM.BIN file stored on the SD-card is not executable code and cannot be loaded
back to the instrument. This coded with md5 algorithm version is used by the authorised verifiers to
check the authenticity of the program installed on the device.
10.5.
Status of “welmec memory” – Status
The
Status
sub-list enables the user to view
the status of the memory dedicated for the
“welmec records” – data and event records
number recorded and available.
.
<ENT>
10.6. Modification of Welmec related parameters
– Service
The
Service
sub-list enables the authorized
users (authorised service centres) to modify
Welmec related parameters and create
event records.
The
Service
sub-list consists of four
positions, enables the user to set the real
time
(
RTC
)
and
perform
calibration
(
Calibration
) without limitation, as well as
input preamplifier and microphone serial
numbers (
Preamplifier
and
Microphone
).
<ENT>
10.6.1. Setting the real-time clock
– RTC
The
RTC
window is a copy of the one
located in the menu
Instrument
. The
difference is that in this place the user may
change time, date and time zone without
any limitations.
<ENT>