11. 1/1 AND 1/3 OCTAVE ANALYSER
The instrument operates as a real time
1/1 Octave
or
1/3 Octave
analyser (RTA) in a very similar way to the
Level Meter
. Moreover,
1/1 Octave
or
1/3 Octave
analysis is performed in parallel with the
SLM
/
VLM
operations. All 1/1-octave (with 10/15 centre frequencies from 16kHz down to 31.5Hz/1.0Hz; in base two system
for Sound/Vibration measurements) and 1/3-octave (with 31/45 centre frequencies from 20kHz down to
20Hz/0,8Hz; in base two system for Sound/Vibration measurements) digital pass-band filters are working in real-
time with the weighting filters (
Z
,
A
,
B
or
C
-
in case of sound analysis;
HP
or
HPE
- in case of vibration analysis),
selected in the
Spectrum
window (
path: Menu / Measurement / Spectrum / Filter
), and the linear RMS (Leq)
detector. This enables the user to pre-weight a spectrum with one of the selected broadband frequency curves
if required for an application such as the provision of hearing protectors in the control of high workplace noise
levels.
Note: The TOTAL RMS results are measured with the weighting filters (A, C and Z - in case of
sound measurements; HP, HP3 and HP10
– in case of vibration measurements) regardless of what
settings were selected in the profiles for Level Meter calculations. The spectra are always linearly
averaged. Thus, the Total values for 1/1 Octave or 1/3 Octave analysis can be different from those
obtained for the profiles (if the RMS Integration was set as Exp).
The
SVAN 977W
instrument operates in two ranges, called
Low
and
High
, which can be selected in the
Range
window
(path: <Menu> / Measurement / Range)
.
The results of
1/1 Octave
and
1/3 Octave
analysis (spectra) can be examined by the
user on a display in the
Spectrum
view.
The availability of this view can be switched
on or off by the user (
path: <Menu> /
Display / Display Modes)
.
The
1/1 Octave
and
1/3 Octave
spectra
are presented for all centre frequencies of
pass-band filters together with the
Total
overall values measured with preselected
frequency
weighting filters. The read-out of
the spectrum value can be done using a
vertical cursor.
11.1. Selection of the 1/1 Octave or 1/3 Octave functions
To select the
1/1 Octave
or
1/3 Octave
analysis function the user should enter the
Function
list by pressing the
<Menu>
push-button, then select the
Function
text
and
press
<ENTER>
.
In
the
Measurement Function
window select
required
(
1/1 Octave
or
1/3 Octave
)
position
and press
<ENTER>
.
<ENT>
Note: It is not possible to change the current function during a measurement.
In this case the instrument displays for about 2 seconds the text:
“Measurement in Progress”.
To change the current function, the measurement must be stopped!