Accelerometer
Optionally general purpose vibration accelerometer 100 mV/g (10 mV/ms
-2
)
(
SV 80
) with a coiled cable for accelerometer 2 m (
SC 27
) can be used.
The SV 80 is an industry standard IEPE accelerometer offered to SVAN 977W.
It is an ideal choice for walk-around vibrations measurement in the rugged
environments of industrial machinery monitoring, such as pumps, motors or
fans. The accelerometer is mounted on a vibrating surface with the mounting
magnet.
The design of SV 80 features the low electronic noise and wide temperature
operating range.
The connection of the accelerometer to the
instrument
is
like
the
microphone
preamplifier, but there is only one ring to be
tighten to the first screw thread.
14.4. Resetting the instrument
SYSTEM RESET:
internal software reset clears any setup configuration, and brings back the default
factory settings. See
Factory Settings
(
path: <Menu> / Auxiliary Setup
).
HARDWARE RESET:
internal hardware reset, no user data is changed. Make sure the battery is not
exhausted, and the unit is turned off. Hold down the
<Alt>
and
<Start/Stop>
push-buttons for more
than 15 seconds, and then release it. Turn on the instrument as usually.
Note: Hardware reset is only to be used in extreme situations such as an instrument hang-up.
Be aware, that a hardware reset:
- will stop any pre-programmed auto-run modes,
- will stop measurement run!
14.5. Firmware upgrade
SVANTEK is committed to continuous innovation path of development, and as such reserves the right to
provide firmware enhancements based on user’s feedback.
To update the instrument firmware:
Unpack the provided firmware package. (provided as a suitable compressed file).
Make sure the unit is turned off.
Connect SC 16 cable to the computer and SV 977W instrument (USB interface).
Keeping pressed the
<ENTER>
and
<
>
push-buttons switch on the instrument - the following
message should appear on the unit's screen: BOOTSTRAP v1.06 (or higher).
Wait for the message
<USB>
on the unit’s screen and start from the PC:
go-usb.bat
.
The changing number and final message: "..... o.k." should appear on the computer screen.
Successful firmware update will be indicated by the message:
Program loaded!
Switch off the instrument.
Note: With the use of Supervisor or + software it is very easy to check if there are any
new firmware releases available for download.