SVAN 973 User Manual
16
The following control keys are located on the front panel of the
instrument:
•
<ESC>
•
<Enter>
•
▲
,
◄
,
►
,
▼
•
<Shift>
•
<Start/Stop>
The action given in (...) brackets denotes the second key function
which is available after pressing it in conjunction (or in sequence)
with the
<Shift>
key.
<Shift>
The second function of a key can be used when the
<Shift>
key is pressed with
<ESC>
or arrow keys. This key can be used in two different modes, which can be
configured in the
Keyboard
screen (
path: <Menu> / Instrument / Keyboard
):
•
like in a computer keyboard, when both
<Shift>
and the second key must be
pressed simultaneously (
Direct
mode);
•
like in a smartphone keyboard, when the first
<Shift>
key should be pressed
and released and then the second key pressed (
2nd Function
mode).
Note: Simultaneous pressing of the <Shift> and <Start/Stop> keys turning the instrument on
or off.
<Start/Stop>
This key allows you to start and stop a measurement process.
<Enter>
This key allows you to open the selected position on the Menu list, to confirm selected
settings or to switch the sub-views in the measurement mode. Some additional
functions of this key will be described in the following chapters of this manual.
<ESC>
This key closes lists of parameters or other screens and return to the upper list of the
menu. It acts in an opposite way to the
<Enter>
key. When a screen is closed after
pressing the
<ESC>
key, any changes just made are ignored.
In the Measurement mode, this key switches results presentation views.
<Shift>+<ESC>
This combination of keys allows you to pause or break the measurement process
temporarily. If there is no current running measurement in progress this key opens the
Setup Manager menu.
◄
/
►
These keys allow you, in particular, to:
•
change viewed result in the measurement mode,
•
select column in a multi-column parameter list,
•
select parameter value in an active position (e.g. filter
Z
,
A
,
C
; integration period:
1s
,
2s
,
3s
, … etc.),
•
control cursor in the
Logger
,
Statistics
and
Spectrum
views,
•
select position of a character in the text editor screen,
•
speed up changing of numerical values of the parameters when pressed and held.
(◄
/
►)
These
◄
/
►
keys pressed in together with
<Shift>
allow you, in particular, to:
•
select parameter value in an active position (e.g. filter
Z
,
A
,
C
;
integration period:
1s
,
2s
,
3s
, … etc.),
•
shift cursor from the first to the last position and back on the plot.
▲
/
▼
These
keys allow you, in particular, to:
•
select position on the list,
•
select correct character in the text editor screen.
(▲
/
▼)
These
▲
/
▼
keys pressed with
<Shift>
allows you, in particular, to:
•
change view in the measurement mode,
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...