SVAN 973 User Manual
40
Note: The Wave Recording function is optional and should be unlocked by entering the
activation code in the text editor screen, opened by the ► key. Once unlocked this option will
be available permanently.
Slope trigger
The
Slope+
trigger starts a signal recording under the condition: rising value of
the Leq result (
Source
) integrated by 0.5 ms passes above the threshold level
(
Level
).
After pressing the
<Start>
key the instrument checks the trigger condition with steps, defined by the
Tr. Period
parameter, and if condition is met starts the signal recording. Recording lasts minimum
Rec. Time
, and during this time the instrument continues to check the trigger condition with the
Tr. Period
step. Provided that
Tr. Period
is shorter than
Rec. Time
, if next trigger condition is met during
Rec. Time
the instrument triggers recording again, so it will continue from this moment by additional
Rec. Time
and so on. If during next recording time there are no triggers, recording will be stopped after
the last trigger plus
Rec. Time
. Assuming, that after first recording trigger conditions continue to be
checked, new wave recording may start during the same measurement time.
The attached example shows that between
measurement start and stop two records
were created. The first record is equal to the
Rec. Time
, because during this period no
second trigger condition has been met.
During
the
second
recording
the
measurement was stopped, and the record
is shorter than
Rec. Time
.
The
Slope -
trigger starts a signal recording under the condition: falling value of
the Leq result (
Source
) integrated by 0.5 ms passes below the threshold level
(
Level
).
This is a mirrored trigger to the
Slope +
trigger.
Note: When a signal recording is waiting for the slope
trigger the “slope”
icon superimposes on the grey „signal” icon.
/
Level trigger
The
Level +
/
Level -
trigger starts a signal recording which will last the
Rec. Time
under the condition: value of the Leq result (
Source
) integrated by 0.5 ms is
greater/lower than the threshold value (
Level
). In other cases, recording doesn’t
start, but if it has been already started it can be continued until the
Rec. Time
has
elapsed.
If during
Rec. Time
a trigger condition
appears, recording will be prolonged for
another
Rec. Time
from the moment of that
trigger condition and so on.
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...