SVAN 973 User Manual
– Appendixes
210
Fig 4. The RT 30
evaluation
E.3.
D
ESCRIPTION OF THE DECAY CURVE RECORDING IN DIFFERENT MEASUREMENT METHODS
➢
DECAY method
This RT 60 measurement method requires omnidirectional sound source which emits pink noise in
appropriate frequency band. The most critical parameter of the omnidirectional sound source is emitted sound
pressure level as it was mentioned in the beginning of the appendix.
The graphical illustration of the data recording in this method is presented in Figure 5.
Fig 5. Data recording in the decay method of the reverberation time evaluation
t [s]
L
[dB]
t
4
t
1
L
1
L
4
5 dB
noise
margin
linear approximation
t
2
t
3
30 dB
t [s]
L
[dB]
t
e
t
s
L
2
t
trig
10 dB
L
1
registration time
t
off
t
on
L
2
L
3
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...