SVAN 973 User Manual
61
9 PRINTING REPORTS
– Report
The
Report
section enables configuring printed reports
of the sound measurement results in the predefined
format.
To open the
Report
section, press the
<Menu>
key,
select the
Report
position and press
<Enter>
.
<ENT>
The
Report
section contains following positions:
allowing you to print measurement results on the default printer;
Options
allowing you to define report options;
Results
allowing you to select measurement results to be included in the report;
Statistics
allowing you to select statistics to be included in the report;
Spectrum
allowing you to select 1/3 octave bands to be included in the report;
Printer
allowing you to select the number of characters in the line of the report.
To obtain the report, connect the instrument to t
he printer’s RS 232 port using the
SV 75
RS 232
interface. This hardware interface is hidden in the Cannon type, 9-pin RS 232 plug-in. On the other end
of the SV 75 interface, which itself looks like a cable, there is the USB-C plug-in. This plug-in should be
placed in the USB-C socket of the instrument.
Be sure that the
RS232
port is properly configured (
path:
<Menu> / Instrument / RS232
). Select in the
RS232
list
the transmission speed (
Baud Rate
) and set the time
limit during which the data transmission should be
performed (
Time Out
).
<ENT>
The RS 232 interface transmission speed (
Baud Rate
) can be selected from the following available
values:
1200
,
2400
,
4800
,
9600
,
19200
,
38000
,
57600
or
115200
bits/s. The transmission speed should
correspond to the same one selected in the printer. The other RS 232 transmission parameters are fixed
to
8 bits for data
,
No parity
&
1 Stop bit
. The default value of the
Time Out
period is equal to one
second, but it can be too short for printers, which are not fast enough. In such cases this parameter
should be increased.
Printers, which have only the USB interface, are currently not driven by the instrument.
Note: Switch the power off before connecting the instrument to any external device (e.g. a
printer or a PC).
Note: All reports are printed in the character format using the ASCII set on either A4 or A5
size paper.
9.1 P
RINTING MEASUREMENT RESULTS
–
P
RINT
The
position enables printing a report on the attached printer or PC.
After pressing
<Enter>
, the instrument checks its current state. If the
measurement is running, printing is not possible, and the appropriate message is
displayed.
If the measurement has been already performed and results are available, the
data will be transferred from the instrument to the attached printer. The instrument
returns to the
Report
list after transferring all data.
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...