SVAN 973 User Manual
23
2.7 O
VERLOAD AND UNDERRANGE DETECTION
Overload detector
The instrument has the built-in overload detectors. Both A/D converter and input amplifier overload
conditions are detected. The overload in the measurement channel (in its analogue part) and the overload
of the analogue / digital converter are both detected.
The “overload” indication appears when the input
signal amplitude is 0.5 dB above
the declared “Peak measurement range”. This condition is checked
once per second or with the Logger Step if it is less than 1 second.
An overload is indicating by the flashing
icon which is displayed during the period from the overload
detection till the end of the Integration Period. If the overload disappears to the Integration Period end,
the overload icon will not be displayed from the start of the next measurement cycle.
When an overload is detected the special marker will be recorded to the logger file with the data logging
step.
The overload time is measured by the
OVL
result during the Integration Period and is saved in the logger
file as part of Summary Results.
Underrange detector
The instrument has the built-
in underrange detector. The “underrange” indication appears when the RMS
value for the elapsed time is below the lower linear operating range. This condition is checked once per
second.
An underrange is indicating by the flashing
icon which is displayed during the period of the
underrange detection. When an underrange is detected till the Integration Period, the special marker will
be recorded to the logger file with the Integration Period step. If during the Integration Period the signal
level increases and the total RMS is greater than the minimum, the icon stops displaying and the
underrange marker is not recording.
2.8 S
AVING DATA
The instrument creates files of the next types:
•
Logger files with
measurement results (extension
.SVL
)
•
Wave files with signal recording (extension
.WAV
)
•
Setup files with measurement and instrument configuration (extension
.SVT
)
Detailed description of structures of all file types is given in Appendix B.
Memory type
All files are stored
in the instrument’s memory in the predefined or assigned directories. The setup files
are stored in the predefined directory SETUP. The non-predefined directories can be changed by the
user or renamed.
File manager
The
File Manager
is used for checking content of the
memory and operations on files and directories such as:
renaming, deleting, displaying information and creating
new directories.
<ENT>
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...