SVAN 973 User Manual
– Appendixes
166
Warm-up time
1 min. (for 0.1 dB accuracy)
Typical stabilization time
after change
in environmental conditions by 20
C
1 hour
Time shift after completion
of a
measurement, before a measurement is
shown
< 1 sec
Note: When the instrument is moved from a warm environment with high humidity,
to a colder environment, care should be taken not to produce condensation inside
the instrument. In this case, much longer stabilization periods may be necessary.
Environmental, electrostatic
and radio frequency
criteria
Effect of humidity
< 0.5 dB (for 30%<RH<90% at 40
C and 1000 Hz)
Effect of magnetic field
below electrical noise level (for 80 A/m and 50 Hz)
Effect of radio frequency fields
meets requirements of IEC 61672-1:2013
The greatest susceptibility (the least immunity) is achieved when in the SLM the
Z
filter and time weighting
F
are selected, and the SPL measurements are considered.
The greatest susceptibility is achieved when the SLM is placed parallel to the radio frequency field.
In addition, if there is an extension cable, the greatest susceptibility is achieved when the SLM and cable is
placed along field and the cable is coil as solenoid.
Effect of electrostatic discharge
meets requirements of IEC 61672-1:2013
During electrostatic discharge, the influence of the displayed results could be observed.
No changes in instrument operation state, configuration or stored data corruption were found out.
Effect of ambient pressure
< 0.01 dB/kPa
Effect of temperature
< 0.5 dB (from -10
C to + 50
C)
Operating
from -10
C to + 50
C
Storage
from -20
C to + 60
C
Microphone
ST 973
free-field condenser microphone
MEMS
(
½” housing)
Nominal sensitivity
1.8 mV/Pa (corresponding to -55 dBV/Pa re 1 V/Pa)
Output impedance
2 x 177.5
Ohm / 15nF, differential
Note: Maximum level of sound pressure level, which can be affect the microphone without
destruction the microphone: 160 dB.
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...