Note:
When logging is waiting for the level trigger the “level” icon
appears alternatively with the
„logger” icon.
/
Source result
Only one measured result (
Source
) can be used for checking trigger condition in the
Level Meter
mode,
namely the instantaneous LEQ from the first profile (with appropriate filter and detector), which is denoted
here as
Leq(1)
. This position cannot be changed.
Threshold level
The threshold value (
Level
) can be set in the range from
24 dB
to
136 dB
. The
Source
value compares with the
Level
value every 0.5 milliseconds.
Pre and post trigger logging
In the
Pre Trigger
position, you can define the number of
results which will be registered in the logger file before
the fulfilment of the triggering condition. This number can
be set in the range 0..10.
In the
Post Trigger
position, you can define the number
of results which will be registered in the logger file after
the fulfilment of the triggering condition. This number can
be set in the range 0..200.
►
These parameters can perform double role. Firstly, when you wish to collect data right after or before the
event that caused logger trigger. Secondly, when it is necessary to have continuous logging, but the
source is oscillating near the threshold level. The extension of the registration window allows to avoid the
effect of pulsation.
Periods of logging before or after fulfilment of the trigger condition are shown to the right of the number
in minutes and seconds (in the format
0m00s
) as a result of multiplication of number of results by the
Logger Step
.
4.5.4 Configuring signal recording
– Wave Recording
The
Wave Recording
position enables activating and
configuring a waveform signal recording in the WAV type
file.
<ENT>
Note: In the Simple interface mode, the Wave Recording position is hidden, but the
instrument will use settings previously defined in the Advanced mode or default settings
(Recording: Off)
.
Note: The Wave Recording function is optional and should be unlocked by entering the
activation code in the text editor screen, opened by the ► key. Once unlocked this option will
be available permanently.
The
Recording
position, if it is not
Off
, defines the way a
signal recording should be done, continuously during
measurement (
Continuous
) or on trigger:
Slope +
,
Slope -
,
Level +
,
Level -
,
Gr
,
Trig.manual
or
Integr. Per
. Default mode:
Off
.
The
File Name
position enables editing the name of the
WAV file.
►..►
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...