SVAN 973 User Manual
69
11 DOSIMETER
The instrument operates as a
Dosimeter
in a very similar way to the
Level Meter
and, in addition to
SLM results, measures also basic dose parameters. This chapter describes Dosimeter specific settings.
Note: One important difference is that in the SEM mode the instrument works in the different
dynamic range than in the SLM mode (see Appendix C)
.
11.1 S
ELECTING
D
OSIMETER FUNCTION
To change the SLM mode to the SEM mode you should
activate the
Dosimeter
function.
To activate the
Dosimeter
function, enter the
Function
section, select the
Meas.
Funct
position and press the
<Enter>
key. In the
Meas. Funct
screen, select the
Dosimeter
function and press the
<Enter>
key.
<ENT>
11.2
S
ETTING GENERAL PARAMETERS
–
G
ENERAL
S
ETTINGS
Most general settings of the
Dosimeter
function are similar to the
Level Meter
). Additionally,
Dosimeter
has a programmable five
automatic pauses.
Programable automatic pauses
Automatic pause(s) can be switched off (
Off
) or can be programmed based on
the RTC time (
On
).
If
Pause
is
On
, two additional positions appear which
enable setting time for pause begin (
Pause
Begin
) and
time for pause end (
Pause End
).
▼
Note: In the Simple interface mode, the Pause parameters are hidden, but the instrument
will use settings previously defined in the Advanced mode or default settings (Off)
.
11.3 S
ETTING PROFILE PARAMETERS
–
P
ROFILE
X
Parameters of three profiles can be set in the
Profile x
lists of parameters.
The following parameters can be programmed
independently for each profile: weighting filter
(
Filter
),
peak filter (
Filter Peak
) and LEQ detector type
(
Detector
), criterion level (
Crit. Level
), threshold level
(Thr. Level
), exchange rate (
Exch. Rate)
, thresholds -
ULT Thresh.
and
PTC Thresh.
.
<ENT>
Weighting filter selection
Z
according to IEC 61672-1:2013 for Class 2,
A
according to IEC 651 and IEC 61672-1:2013 for Class 2,
C
according to IEC 651 and IEC 61672-1:2013 for Class 2,
B
according to IEC 651 for Class 2.
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...