5) Hang weight platform and 10% weights on the calibration arm, gently. Wait until the
arm has completely stopped moving.
6) The torque displayed on the LCD's should be within the accuracy tolerance for the 10%
calibration point for the unit.
7) Repeat steps 5 and 6 with the 20%, 50%, 80%, and 100% loads as specified in the
"
Calibration Table -
Torq-Tronics
® Models
" at the end of this chapter.
8) Remove and reverse calibration arm to the counter-clockwise direction, and press the
Zero key.
9) Repeat steps 3, 4 and 5 in the counter-clockwise direction.