Low Frequency Leakage Current Test
– Argon Active
1. Reference
Figure 13-4
below.
2. Enter the
“Patient Leakage Measurement Mode” on the ESA.
3. Connect the 4-mm test lead to the right port of the ArConnect® Receptacle on the front
panel of the gi4000 ESU.
4. Connect the test lead to the patient lead receptacle on the ESA.
5. Measure low frequency (LF) leakage current between the active portion of the argon
receptacle on the gi4000 ESU and the ESA in the following two (2) test conditions:
Normal Condition Test
- NORMAL POLARITY, CLOSED NEUTRAL, OPEN EARTH.
Single Fault Condition Test
- REVERSE POLARITY, CLOSED NEUTRAL, OPEN EARTH
6. Record the measured value from the ESA on the 00732974 Preventative Maintenance
Test Data Summary Sheet
– gi4000 ESU.
7. Unplug the test lead from the ArConnect® Receptacle.
Power Cord
Figure 13-4: LF leakage Current Test Diagram
– Argon Active