00732467 Rev. D
Page
80
of
110
Table 12-1 (Cont
’d)
Text On Screen
Alarm Priority
Action
FAULT:
FPGA argon bottle high
pressure. Contact
Technical Services.
High Priority
Contact STERIS Customer Service.
FAULT:
FPGA high voltage
relay check. Contact
Technical Services.
High Priority
Contact STERIS Customer Service.
FAULT:
Temperature failure.
Turn off and
let cool down.
High Priority
Power OFF. Let unit cool down before turning unit
ON.
FAULT:
Touchscreen circuit failure.
Contact Technical Services.
High Priority
Do not touch the Touchscreen during POST. Power
OFF/ON. If the fault does not reappear, the gi4000
is OK. If POST is not successful or if the fault
screen returns, then Contact STERIS Customer
Service.
FAULT:
SLCD macro file
version failure. Contact
Technical Services
High Priority
Contact STERIS Customer Service.
FAULT:
SLCD communications
failure. Contact Technical
Services.
High Priority
Contact STERIS Customer Service.
FAULT:
Disconnect Stuck
Footswitch
High Priority
Footswitch may be sticky or dirty. Soak and clean.
Do NOT submerse Footswitch connector.
FAULT:
Footswitch Circuit Failure
Contact Technical Services
High Priority
Contact STERIS Customer Service.
FAULT:
Pad circuit failure.
Contact Technical Services.
High Priority
Contact STERIS Customer Service.
FAULT:
Relay failure.
Contact Technical Services.
High Priority
Contact STERIS Customer Service.
FAULT:
RF output failure.
Contact Technical Services.
High Priority
Power OFF/ON. If the fault returns, contact STERIS
Customer Service.
Содержание G1110001
Страница 1: ...gi4000 Electrosurgical Unit Reference Reorder No G1110001 User and Maintenance Manual ...
Страница 6: ... This page is intentionally left blank ...
Страница 54: ...00732467 Rev D Page 48 of 110 8 11 9 Bipolar Method Figure 8 25 Figure 8 26 Figure 8 27 ...
Страница 109: ...00732467 Rev D Page 103 of 110 Figure 13 21 Argon RF Output Test Diagram ...