8.9 Duty Cycle
Generator power is continuously ON when the unit is powered ON. All modes under maximum
power settings and rated load conditions (monopolar modes: 500
Ω, bipolar mode: 100Ω) have
an output activation time of 10 sec ON / 30 sec OFF.
8.10 Leakage
Low Frequency (50-60 Hz) Leakage Current
Chassis to Neutral
Line Polarity to Nominal (Ground Open):
< 150
A
Line Polarity Reversed (Ground Open):
< 150
A
Patient Leads to Neutral (All RF Active and Patient Plate)
Line Polarity to Nominal (Ground Open):
< 10
A
Line Polarity Reversed (Ground Open):
< 150
A
Radio Frequency (RF) Leakage Current
Monopolar Method (All Modes at Maximum Power Setting)
Patient Plate to Ground:
< 150
A
Active to Ground:
< 150
A
Bipolar Method (Maximum Power Setting)
Each Lead to Ground:
< 54
A
8.11 Output Power vs. Resistance Graphs
The below graphs show characteristics for “Power vs. Load Resistance”, “Power vs. Power
Setting” and “Open Circuit Peak Voltage vs. Power Setting” for each mode of operation. The
test setup for these graphs is described in IEC 60601-2-2. Variation of this test setup will vary
the measured values.