00732467 Rev. D
Page
105
of
110
Figure 13-24
6. Remove old filter by turning luer lock connection counterclockwise.
7. Install new filter by threading luer fittings into filter ends and secure by turning clockwise.
Ensure that the molded text on the filter is facing the IN label on the argon tubing, as
shown in
Figure 13-25
.
Figure 13-25
8. Dispose of used filter in accordance with your facility procedures and policies.