00732467 Rev. D
Page
64
of
110
10.1.2 Lavage Function
1.
Engage Standby Mode by pressing the standby button on the touchscreen display
or gi4000 footswitch.
2.
Open the bottle of sterile water and remove the cap. The water bottle holder
provided with the gi4000 ESU is compatible with most disposable sterile water
bottles.
3.
Before installing irrigation tubing:
•
Visually inspect the irrigation tubing for damage (holes, tears, contamination,
etc.).
•
Verify the pump head is adjusted to the appropriate irrigation tubing setting.
Reference section
9.5 System Installation and Setup: Setup
– Lavage
Pump
for instructions on adjusting the pump head tubing adjusters.
4.
Connect irrigation tubing to sterile water bottle per manufacturer’s instructions.
5.
Open the pump head lid of the lavage pump.
6.
Locate the positional markers on the irrigation tubing and insert into the pump head
between the positional markers.
7.
While placing the irrigation tubing into the pump head, grasp the tubing together and
gently pull upward, as shown in
Figure 10-1.
This will ensure that the irrigation tubing
is centered and aligned properly in the Pump Head to avoid kinking, tearing, or
disruption of water flow.
8.
Ensure the irrigation tubing is installed into the pump head
in the direction of the
arrow. Water flow is right to left when facing the pump.
9.
Close the pump head lid. Verify the irrigation tubing is not pinched in the pump head
lid.
10. Prior to insertion of the endoscope into the patient, prime the irrigation tubing and
the water irrigation channel/accessory channel of the endoscope by depressing the
irrigation pedal on the footswitch.
NOTE:
During use of the gi4000 ESU, the user should be within three to five
feet of the unit to interact with the touchscreen display and unit features.
Figure 10-1
Содержание G1110001
Страница 1: ...gi4000 Electrosurgical Unit Reference Reorder No G1110001 User and Maintenance Manual ...
Страница 6: ... This page is intentionally left blank ...
Страница 54: ...00732467 Rev D Page 48 of 110 8 11 9 Bipolar Method Figure 8 25 Figure 8 26 Figure 8 27 ...
Страница 109: ...00732467 Rev D Page 103 of 110 Figure 13 21 Argon RF Output Test Diagram ...