00732467 Rev. D
Page
95
of
110
Radiofrequency Leakage Current Test
– Monopolar (COAG, Blend COAG, Blend CUT, CUT)
1. Select the Monopolar Method on the gi4000 ESU touchscreen display.
2. Reference
Figure 13-12
below.
3. Connect the 8-mm to 4-mm monopolar test adaptor to the Monopolar Receptacle on
the front panel of the gi4000 ESU.
4. Connect the active test lead to the monopolar test adapter on the front panel of the
gi4000 ESU.
5. Connect the monopolar test lead to the active input receptacle on the ESU Analyzer.
6. Reference
Figure 13-13
below.
7. Connect the earth ground test lead to the ground terminal on the rear panel of the
gi4000 ESU.
8. Connect the earth ground lead to the earth ground input receptacle on the ESU
Analyzer.
9. Connect a dummy dispersive electrode test lead to the PSS Receptacle on the gi4000
ESU (to receive yellow indicator), as shown in
Figure 13-12
below.
10. Measure the RF leakage current from both the active and dispersive sides for each
Monopolar Mode (COAG, Blend COAG, Blend CUT, CUT).
11. Record the measured value(s) from the ESU Analyzer on the 00732974 Preventative
Maintenance Test Data Summary Sheet
– gi4000 ESU.
Figure 13-12: RF Leakage Measurement Test Diagram
– Monopolar Active and Dispersive
Содержание G1110001
Страница 1: ...gi4000 Electrosurgical Unit Reference Reorder No G1110001 User and Maintenance Manual ...
Страница 6: ... This page is intentionally left blank ...
Страница 54: ...00732467 Rev D Page 48 of 110 8 11 9 Bipolar Method Figure 8 25 Figure 8 26 Figure 8 27 ...
Страница 109: ...00732467 Rev D Page 103 of 110 Figure 13 21 Argon RF Output Test Diagram ...