TUBE REPLACEMENT
continued
1.
Ensure the power to the pump is off and the input
signal is disconnected.
2.
Place the new tube in the pump head and use
your fingers to center it on the rollers.
H
3.
Place the tube housing cover (feet first) on the tube
housing, affix the front of the latches to the cover
lip and then press the latches back to secure. Be
sure the cover is seated with the sleeve bearing on
the shaft and is flush with the housing before
latching.
I
4.
Plug the pump in. With the pump head cover
latched, allow the pump to run the roller assembly
in its collapsed position for one minute by pressing
and holding the PRIME and MODE at the same
time and hold for one minute.
J
5.
Unplug the power cord.
6.
Remove the tube housing cover and flip to use as a
tool in the next step.
K
7.
Align the center of the inverted cover with the
center of the roller assembly so that the three holes
on the face of the cover align with the three knurled
lugs on the roller assembly. Position the cover feet
near the bottom.
L
I
Install cover feet first
H
Place new tube
Cover Feet
L
Align cover feet near bottom
3 Lugs
Center of Roller Assembly
Cover Feet
INSTALL THE TUBE/EXPAND THE ROLLER ASSEMBLY
Cover
J
Place pump in PRIME
K
Invert cover
60 seconds
PRIMe
50%