28
S10P Series www.stenner.com
LEAK DETECT
continued
CALIBRATE THE LEAK DETECT SENSITIVITY
page 2 of 2
7.
Soak a small piece of sponge with the pumping solution and place over the two leak
detect pins, use the expected weakest solution and keep in mind some solutions
dilute with time.
8.
Slowly turn the potentiometer counterclockwise until the display screen shows LEAK.
9.
Turn the potentiometer an additional one full turn counterclockwise.
10.
Remove sponge and thoroughly clean the solution off pins and confirm they are dry.
IMPORTANT:
Confirm there is no chemical residue remaining on the
leak detect components.
11.
Disconnect, then reconnect power. Confirm the display screen does not show LEAK.
If display shows LEAK, repeat steps 1-11. If the display does not show LEAK, go
to step 12.
12.
If the pump is not outdoors or exposed to water, go to step 14.
13.
If the pump will be installed outdoors or exposed to water:
- Soak a small piece of sponge in water
and place over the two leak detect pins.
If the display shows LEAK, it indicates the
conductivity of the pumped solution and
water is too close and the pump cannot
discriminate between the two. The liquid
end needs to be protected from water
intrusion to avoid a false leak detection.
- If the display does not show LEAK, the setting is complete.
14.
Re-install the tube housing cover and the signal cover on the pump.
15.
Prime the pump.
16.
Verify pump operation.
LeAK
60 seconds
Mode