PMX MANUAL
20
118141-002 REV
JB1
Customer Interface
High Voltage
power supply
1
7
12
5
8
+5V
10K
Transistor On =
Rotor shutdown
+5V
10K
Transistor On =
X-ray OFF
+24V
Switch closed
for PREP
+24V
Switch closed
for EXPOSURE
NOTES:
Relay coil Z=178
Ω
+5V
24
+24V
Figure 3-2 Digital I/O Control Interface
3.2.4
Remote Status Interface
Remote status signals for KV MIN tube current
limit, tube power limit, overvoltage, and filament
Current limit, over temperature, and power
supply fault is provided.
These signals are open collector type.
(Normally off), and are intended to drive the
diode of an opto-coupler.
Figure 3-3 for recommended interface circuits.