PMX MANUAL
38
118164-001 REV. A
Table 6 – Recommended separation distances between portable and mobile RF communications equipment
and the ME EQUIPMENT or ME SYSTEM – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-
SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment
and the PMX
The PMX is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the PMX can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the PMX
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output
power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d
= 1.17
√
P
80 MHz to 800 MHz
d
= 1.17
√
P
800 MHz to 2,5 GHz
d
= 2.33
√
P
0.01
0.117
0.117
0.233
0.1
0.370
0.370
0.737
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where
P
is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.