118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
48
4.4.22
Set Brake After Exposure
Description:
The host requests that the firmware set the Brake After Exposure setting.
When Break After Exposure is on, the unit will begin breaking (stopping
the rotor) immediately after the end of the current exposure. In this mode it
is required that the host begin a new Prep cycle (Prep inactive to active)
before a new exposure can be started. When Break After Exposure is off,
the rotor remains running past the end of the exposure for a period of time
specified by the Break after Exposure Hang time. As long as prep remains
active and the Break after Exposure Hang time does not expire, a new
exposure can be started without cycling Prep. See Section 7.0 for more
details on Break After Exposure operation. Note, there is no
corresponding Get Break After Exposure command. The current Break
After Exposure setting is available using the Request Status Command.
Direction:
Host to supply
Syntax:
<STX><35><,><ARG1><,><CSUM><ETX>
Where:
<ARG1> = Brake after exposure, 1 or 0 in ASCII format
1 = Break After Exposure mode is On
0 = Break After Exposure mode if Off
Example:
<STX>35,1,<CSUM><ETX>
Response:
<STX><35><,><$><,><CSUM><ETX>