118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
47
4.4.21
Get Last Exposure Time
Description:
The host requests that the firmware sends the last exposure time
length, as measured from the 75% of programmed value points.
This is the actual exposure time of the last exposure, not the value
set by the “Set Exposure Settings” Command.
Direction:
Host to supply
Syntax:
<STX><65><,><CSUM><ETX>
Example:
<STX>65,<CSUM><ETX>
Response:
<STX><65><,><ARG1><,><CSUM><ETX>
Where:
<ARG1> = last exposure time in milliseconds (in ASCII format)
Example:
<STX>65,4095,<CSUM><ETX>