118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
82
7.0
Brake After Exposure
When ‘BAE’ is disabled, the Brake should happen after the hang timer reaches the set Hang Time
or the Prep is removed whichever occurs first. If Hang Time = 0, the brake will happen once the
exposure command is removed (the same as BAE enabled)..
When ‘BAE’ is enabled, the brake should happen once Exposure command goes inactive.
After brake occurs, the generator is no longer at READY status. Prep sequence has to be initiated
(Prep goes low then high).
Prep
Exposure
Ready
Filament Inhibit
Enable MA Program
Enable KV Program
DSP Inhibit
Xray On
Ma Servo
HSS Low
Speed States
HSS High
Speed States
STOP
STOP
BOOST
BOOST
RUN
STOP
RUN
RUN
Hang
Time
RUN
Hang
Time
BRAKE
STOP
BOOST
BOOST
RUN
STOP
RUN
BRAKE
STOP
Prep then Exposure
Prep & Exposure
Together
Figure 14: BAE Disabled Timing