overwritten if the unit is powered up. 0 = Unit keeps current
FRAM contents (normal operation).
ARG13 = Ready Status, 1 = Unit is ready to make X-Rays, 0 = Unit
is not Ready.
ARG14 = Setup Error, 1= The current exposure setup programmed
into the unit is invalid. The unit will not be allowed to turn on
high voltage, and attempting to do so will result in a “setup
fault”. 0 = The current exposure setup is valid.
Note that this status ARG is not a “fault”. You do not have to
clear the fault register to clear the Setup Error flag, you just
have to correct the setup.
ARG15 = Calibration Mode. 1 = The unit is in calibration mode. It
has received a command to enter calibration mode, and the
calibration is in progress, or waiting for the user to start it.
0 = The unit is not in calibration mode. (See section 11 on
calibration for more information)
ARG16 = Filament Open Loop. 0 = Normal operation (Filament
closed loop). 1 = Test mode - Filament control feedback loop
is disabled.
ARG17 = AC/DC Fault bypass. 0 = Normal operation (line voltage
input fault not bypassed). 1 = Test mode (line voltage fault is
bypassed).
ARG18 = Open Filament Fault bypass, 0 = Normal operation (unit
faults on open filament). 1 = Test mode (unit will not fault if
open filament).
ARG19 = Analog Programming Mode, 0 = Normal operation (unit is
under digital control from serial interface). 1 = Test mode
(unit is programmed via analog control).
ARG20 = Over duty Fault bypass. 0 = Normal operation (over duty
fault not bypassed). 1 = Test mode. Unit will not fault if
maximum duty cycle exceeded.
ARG21 = kV / mA hold Bypass. 0 = Normal operation, kV and mA
feedback values are held for 2 seconds after the end of the
exposure. 1 = Values are not held. (See note 3 for the
analog monitor read back command for more information).
ARG22 = Over Power Bypass. 0 = Normal Operation. 1 = Test
mode over power fault bypassed.
ARG23 = Inverter Over Temperature. 0 = Inverter Temperature OK.
1 = Inverter Temperature Sensor reads over 60c, not OK to
start an exposure. (See note 2 below)
ARG24 = Duty cycle limit ok. 1 = Ok to enter Prep. 0 = Not ready to
enter prep because to do so would put the unit in an Over-
duty condition. This bit indicates if the unit over-duty
protection will allow a new exposure cycle to begin. It
combines the 20 second between exposure, 5 second on
time in 2 minutes, and 2 Rotor starts in 1 minute limitations.