118165-001 REV. A
475 Wireless Boulevard • Hauppauge, New York 11788, USA • www.spellmanhv.com • T:+1 631.630.3000 • F:+1 631.435.1620
45
4.4.19
Request Last Exposure kV Monitor
Description:
The host requests that the firmware report the kV monitor value for
the last exposure. This is the average value of the exposure kV for
the previous exposure. If this value is read during the exposure the
average value to that point is reported.
The last exposure average value will be held until either a
new exposure is started or a new prep cycle is started (Prep goes
from inactive to active). If either of these conditions occurs, the held
value is discarded and the current kV value is reported.
The kV monitor value is from 0 – 4095, corresponding to voltages
0V – 53476V, (10.44V / Bit).
Direction:
Host to supply
Syntax:
<STX><60><,><CSUM><ETX>
Response:
<STX><60><,><ARG><,><CSUM><ETX>
Where:
<ARG> = 0 - 4095 in ASCII format (0-4095 = 0-53476 V)
Scaling = 53476/4095 = 13.06 V / Bit
Example:
<STX>60,4095,<CSUM><ETX>