Service record Dynamis MV
13 von 36
2021-01-12
In preparation, please remove the drive wheels
Remove the screw connection on both sides
(3 A)
of the seat support brackets
(3 B)
and
pull the sleeves in the front screw connections
outwards so that the seat support brackets can
be moved freely.
Put the seat support bracket in position, push
the sleeves back into the selected holes
(3 B)
of the seat support bracket and tighten the
screws on both sides.
3.2.5 Seat height adjustment
(3)
3.2.3 Axis of rotation seat unit/ back unit
The axis of rotation should be aligned as close
as possible to the patient's hip joint. We re-
commend using the outer hole (1 B) for a body
height of up to 150 cm and the inner hole (1 A)
for larger body sizes.
If the screw connection is removed in order to determine a new position, the screw con-
nection must be secured with medium-strength screw locking.
(A)
(B)
3.2
ERGO-seat assembly
3.2.4 Growth in seat depth ERGO seat
The seat depth of the ERGO seat can be adjus-
ted by + 2cm.
To do this, the following steps should be ob-
served:
• Remove the seat plate screws.
(2 A)
• Bring the seat plate into the desired
position and insert the screws in the
respective holes.
(2 B)
• Remove the screws from the seat
center section / side section and
screw them back into the center hole.
(2 C)
• Check all screws to ensure that they are
firmly tightened again.
(A)
(A)
(B)
(B)
(C)
(C)
(1)
(2)
(A)
(B)