M S 1 6 A S y r i n g e D r i v e r
Te c h n i c a l S e r v i c e M a n u a l
I s s u e 6 ( F e b r u a r y 2 0 0 5 )
P a g e v i i
S m i t h s M e d i c a l
Warnings
Warnings tell you about dangerous conditions that may occur, and
which could lead to death or serious injury to the user or patient if
you do not obey all of the instructions in this manual.
1.
Remember the rate is set in milliMETRES
NOT milliLITRES. The correct rate is essential to
prevent serious injury or death to the patient.
2.
Any damaged pumps must be thoroughly tested by a competent person in accordance with the
procedures in this service manual before re-use with a patient. Failure to observe this may result
in patient injury or death.
2.
Completely prime the administration set, and remove all air from both the administration set and
the syringe before measuring the mm of fluid length, otherwise the rate calculation will be
incorrect.
3.
The driver must only be used with the syringe rubber securing strap fixed firmly in place, thus
preventing an uncontrolled infusion to the patient that could result in serious injury or death.
4.
Never take a syringe that is not empty off the driver if it is still connected to the patient.
The infusion line must first be clamped or disconnected to prevent serious injury or death to the
patient.
5.
The use of spare parts other than those recommended is not advised, and may affect the correct
operation of the driver, resulting in injury or death.
6.
The syringe driver must not be used for infusing medication where pulsatile delivery action is
unacceptable.
7.
The syringe driver must not be used in environments with flammable gas mixtures or with oxygen-
enriched atmospheres.
8.
If the syringe driver is used in the lockbox, the alarm volume will be reduced.
SM0105-CON-6.PMD
2/23/2005, 12:12 PM
7
Содержание Graseby MS16A
Страница 13: ...1 Graseby MS16A Syringe Driver Chapter 1 Product Overview SM0105 1 6 PMD 2 23 2005 2 54 PM 1 ...
Страница 18: ...2 Graseby MS16A Syringe Driver Chapter 2 Specification and Standards SM0105 2 6 PMD 2 23 2005 2 55 PM 1 ...
Страница 23: ...3 Graseby MS16A Syringe Driver Chapter 3 Circuit Descriptions SM0105 3 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 29: ...4 Graseby MS16A Syringe Driver Chapter 4 Disassembly and Assaembly SM0105 4 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 40: ...5 Graseby MS16A Syringe Driver Chapter 5 Service Test Proceedures SM0105 5 6 PMD 2 23 2005 3 07 PM 1 ...
Страница 50: ...6 Graseby MS16A Syringe Driver Chapter 6 Maintenance SM0105 6 6 PMD 2 23 2005 3 03 PM 1 ...
Страница 55: ...7 Graseby MS16A Syringe Driver Chapter 7 Parts List SM0105 7 6 PMD 2 23 2005 2 59 PM 1 ...