Te c h n i c a l S e r v i c e M a n u a l
I s s u e 6 ( February 2 0 0 5 )
5 — 3
S m i t h s M e d i c a l
M S 1 6 A S y r i n g e D r i v e r
5
Chapter 5 - Service Test Procedures
This chapter describes the functional testing of the Syringe Driver.
The routines must be followed whenever the case is opened,
components replaced or any repair completed
Typical situations include:-
●
routine planned preventative maintenance (PPM),
●
after the pump has been opened
whether or not repaired
,
●
whenever it is required to verify that the pump is safe to use.
N
ote
:
The case must be resealed if the case halves have been
separated.
Smiths Medical recommends that the Syringe Drivers are
functionally and safety tested annually, particularly if they’ve been
accidently maltreated.
Although the case is sealed, it is not intended to be fully waterproof.
It is merely a matter of good practice to check the functionality of a
unit before using it in a patient-critical situation.
Tools and equipment
The following tools and equipment are required to complete the
adjustments, tests and calibrations:
Description
Part number
Cam adjustment tool
0105-0079
Thrust rig
0105-0651
Spring test gauge assembly or weights:
0105-0083
1.5 kg, 2 kg, 3 x 5 kg
Stop watch - 1/2 sec resolution
Linear Accuracy Rig
0105-0858
Vernier calipers
0 - 10 V variable DC supply at up to 50 mA.
*Dummy rear case (refer to Smiths Medical Service Dept.)
Standard Service Department tools and equipment.
* A dummy rear case consists of a rear case that has been cut in
half so that only the battery end of the case (with two fixing
holes) remains.
SM0105-5-6.PMD
2/23/2005, 3:07 PM
3
Содержание Graseby MS16A
Страница 13: ...1 Graseby MS16A Syringe Driver Chapter 1 Product Overview SM0105 1 6 PMD 2 23 2005 2 54 PM 1 ...
Страница 18: ...2 Graseby MS16A Syringe Driver Chapter 2 Specification and Standards SM0105 2 6 PMD 2 23 2005 2 55 PM 1 ...
Страница 23: ...3 Graseby MS16A Syringe Driver Chapter 3 Circuit Descriptions SM0105 3 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 29: ...4 Graseby MS16A Syringe Driver Chapter 4 Disassembly and Assaembly SM0105 4 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 40: ...5 Graseby MS16A Syringe Driver Chapter 5 Service Test Proceedures SM0105 5 6 PMD 2 23 2005 3 07 PM 1 ...
Страница 50: ...6 Graseby MS16A Syringe Driver Chapter 6 Maintenance SM0105 6 6 PMD 2 23 2005 3 03 PM 1 ...
Страница 55: ...7 Graseby MS16A Syringe Driver Chapter 7 Parts List SM0105 7 6 PMD 2 23 2005 2 59 PM 1 ...