Te c h n i c a l S e r v i c e M a n u a l
I s s u e 6 ( February 2 0 0 5 )
5 — 9
S m i t h s M e d i c a l
M S 1 6 A S y r i n g e D r i v e r
5
6.
Timer and feed rate tests
Place the syringe driver in the Linear Accuracy Rig,
part number
0105-0858
, place the dti extension rod on the actuator assembly and
set the dti to zero.
a.
With a supply of 9 volts, set the rate to 55 and check that the
time for 5 motor pulses to occur is 36 - 41 secs. Ensure that the
actuator advances along the leadscrew a distance of 0.55 - 0.61 mm
in this time.
b.
Set the rate to 66 and check that the time for 5 motor pulses to
occur is 30 - 33.5 secs.
c.
Set the rate to 99 and check that the time for 10 motor pulses to
occur is 40 - 45 secs.
Hold down the
START/TEST
pushbutton
and check that the motor starts and then stops and sounder
cuts in between 3.3 - 4.0 secs.
d. Release the
START/TEST
pushbutton. Observe that the unit
restarts, with flashing LED and a single motor pulse.
7.
Cam alignment
Increase power supply to 10 V DC and check that six pulses equals
one revolution of the leadscrew.
8.
Half nut slip
Set the supply at 10 V DC and rate at 99. Check for correct stalling.
The LED must stop flashing, the alarm must sound, and the half nut
must
not
slip, when the actuator runs to the end of the leadscrew.
Repeat at least twice.
9.
Rate switch test
a.
Set rate to 36 and check that the time for 10 motor pulses to
occur is 11 - 13 secs.
b.
Set rate to 00 and start the driver (LED flashing). Check that no
motor pulses occur over a period of at least 8 minutes. This can
be most easily determined by an absence of actuator movement.
The syringe driver may be removed from the Linear Accuracy Rig
before continuing.
10. Current tests
a.
At a supply voltage of 9 V DC, measure the supply current with
the unit in the OFF state (LED not flashing, alarm silent). Check
this does not exceed 5 mA.
b.
Set rate to 00 and start unit (LED flashing). Check that the
supply current (excluding LED flashes) does not exceed 50 mA at
any time in a 30 second test period.
SM0105-5-6.PMD
2/23/2005, 3:07 PM
9
Содержание Graseby MS16A
Страница 13: ...1 Graseby MS16A Syringe Driver Chapter 1 Product Overview SM0105 1 6 PMD 2 23 2005 2 54 PM 1 ...
Страница 18: ...2 Graseby MS16A Syringe Driver Chapter 2 Specification and Standards SM0105 2 6 PMD 2 23 2005 2 55 PM 1 ...
Страница 23: ...3 Graseby MS16A Syringe Driver Chapter 3 Circuit Descriptions SM0105 3 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 29: ...4 Graseby MS16A Syringe Driver Chapter 4 Disassembly and Assaembly SM0105 4 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 40: ...5 Graseby MS16A Syringe Driver Chapter 5 Service Test Proceedures SM0105 5 6 PMD 2 23 2005 3 07 PM 1 ...
Страница 50: ...6 Graseby MS16A Syringe Driver Chapter 6 Maintenance SM0105 6 6 PMD 2 23 2005 3 03 PM 1 ...
Страница 55: ...7 Graseby MS16A Syringe Driver Chapter 7 Parts List SM0105 7 6 PMD 2 23 2005 2 59 PM 1 ...