I s s u e 6 ( F e b r u a r y 2 0 0 5 )
Technical Ser vice Manual
S m i t h s M e d i c a l
4 — 6
4
M S 1 6 A S y r i n g e D r i v e r
The
START/TEST
pushbutton, seal and switchplate (items 5, 6
and 7) may remain in the case.
e.
To release the piezo-transducer (see
Figure 4-2
) assembly, unplug
IC1 from its socket, exposing the fixing clip. Remove the piece of
rubber from the clip. Do not pull on the metal disk. Push the clip
through, whilst gently holding the transducer assembly.
Figure 4-2 Release piezo-transducer assembly
f. To remove the PCB assembly completely, unsolder the motor
gearbox assembly (item 14) from the PCB.
g. Ensure that all traces of sealant (if applied) are removed from
mating surfaces.
5. Motor and gearbox disassembly
a.
Separate the case halves and remove the leadscrew and bearing
assembly as described in paragraphs 1, 2 and 3 above.
b.
Remove the PCB assembly (see paragraph 4, a - d above).
c.
Unsolder the motor and gearbox assembly (item 13) from the
PCB.
SM0105-4-6.PMD
2/23/2005, 2:56 PM
6
Содержание Graseby MS16A
Страница 13: ...1 Graseby MS16A Syringe Driver Chapter 1 Product Overview SM0105 1 6 PMD 2 23 2005 2 54 PM 1 ...
Страница 18: ...2 Graseby MS16A Syringe Driver Chapter 2 Specification and Standards SM0105 2 6 PMD 2 23 2005 2 55 PM 1 ...
Страница 23: ...3 Graseby MS16A Syringe Driver Chapter 3 Circuit Descriptions SM0105 3 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 29: ...4 Graseby MS16A Syringe Driver Chapter 4 Disassembly and Assaembly SM0105 4 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 40: ...5 Graseby MS16A Syringe Driver Chapter 5 Service Test Proceedures SM0105 5 6 PMD 2 23 2005 3 07 PM 1 ...
Страница 50: ...6 Graseby MS16A Syringe Driver Chapter 6 Maintenance SM0105 6 6 PMD 2 23 2005 3 03 PM 1 ...
Страница 55: ...7 Graseby MS16A Syringe Driver Chapter 7 Parts List SM0105 7 6 PMD 2 23 2005 2 59 PM 1 ...