I s s u e 6 ( F e b r u a r y 2 0 0 5 )
Te c h n i c a l S e r v i c e M a n u a l
S m i t h s M e d i c a l
1 — 4
M S 1 6 A S y r i n g e D r i v e r
1
The clear plastic lockbox is hinged at the back and features a lock at
the front. The syringe is held in place on the driver by the securing
strap and the infusion line passed through the slot at the front of the
box. When the
mm PER
setting has been selected and the
START/
TEST
pushbutton pressed, the loaded driver is located in the box
which is then locked. The patient controls (
Figure 1-3
) may be viewed
through the clear window at the side of the lockbox.
Controls
Patient controls (
Figure 1-3
) consist of two rotary set rate
mm PER
slotted switches, and a
START/TEST
pushbutton.
START/TEST
50
60mm
GM0418-A
mm PER
9
GM0307-A
0
Figure 1-3 MS16A Controls
Operation
Note
: For detailed operating information, refer to the MS16A, MS26
Instruction Manual,
part number 0105-0549
.
To start an infusion, fit a primed syringe and press the
START/TEST
pushbutton.
When held down, the
START/TEST
pushbutton tests a safety cut-out
circuit. When released, it starts the drive circuit. Once started, the
front panel LED flashes approximately every second.
When the plunger reaches the end of travel (syringe empty) or is
occluded, the driver switches off, the LED stops flashing and an
audible alarm sounds.
If the
START/TEST
pushbutton is held down for more than 3 seconds,
the alarm sounds, the driver stops and the LED stops flashing.
If required, the driver can be modified to sound a continuous tone
when the driver stops. See Chapter 6, Maintenance for modification
instructions.
For additional protection, the syringe driver can be secured in a
Lockbox.
SM0105-1-6.PMD
2/23/2005, 2:54 PM
4
Содержание Graseby MS16A
Страница 13: ...1 Graseby MS16A Syringe Driver Chapter 1 Product Overview SM0105 1 6 PMD 2 23 2005 2 54 PM 1 ...
Страница 18: ...2 Graseby MS16A Syringe Driver Chapter 2 Specification and Standards SM0105 2 6 PMD 2 23 2005 2 55 PM 1 ...
Страница 23: ...3 Graseby MS16A Syringe Driver Chapter 3 Circuit Descriptions SM0105 3 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 29: ...4 Graseby MS16A Syringe Driver Chapter 4 Disassembly and Assaembly SM0105 4 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 40: ...5 Graseby MS16A Syringe Driver Chapter 5 Service Test Proceedures SM0105 5 6 PMD 2 23 2005 3 07 PM 1 ...
Страница 50: ...6 Graseby MS16A Syringe Driver Chapter 6 Maintenance SM0105 6 6 PMD 2 23 2005 3 03 PM 1 ...
Страница 55: ...7 Graseby MS16A Syringe Driver Chapter 7 Parts List SM0105 7 6 PMD 2 23 2005 2 59 PM 1 ...