Te c h n i c a l S e r v i c e M a n u a l
I s s u e 6 (February 2 0 0 5 )
6 — 5
S m i t h s M e d i c a l
M S 1 6 A S y r i n g e D r i v e r
6
Continuous alarm conversion
If required, the alarm may set to a continuous tone when the syringe
stops e.g. at the end of an infusion.
To change the alarm tone:
1.
Separate the case halves and remove the leadscrew and bearing
assembly (see
Chapter 4, Disassembly and Assembly
).
2.
Remove one screw, one fibre washer and the motor clamp (see
Figure 4-1
).
3.
Remove the remaining four screws and fibre washers securing
the PCB assembly to the front case.
If the wires for the piezo-transducer assembly are secured with
an adhesive pad to the motor clamp, carefully lift the leads from
the pad.
4.
Pull the battery connector springs from their locating slots and
lift out the PCB assembly complete with microswitches and
motor and gearbox assembly.
The
START/TEST
pushbutton, seal and switchplate may remain
in the case.
5.
Remove the PCB completely.
6.
Remove capacitor C5 and resistor R17 from the board.
Figure 6-2 Continuous alarm conversion
7.
Fit resistor R17 in the C5 position.
8.
Assemble the driver (see
Chapter 4, Disassembly and Assembly
).
9.
Test the driver (see
Chapter 5, Testing
).
Ensure that the alarm tone is continuous when the motor stops.
10. Check that the alarm stops when an infusion starts (and when
the battery is removed.
SM0105-6-6.PMD
2/23/2005, 3:03 PM
5
Содержание Graseby MS16A
Страница 13: ...1 Graseby MS16A Syringe Driver Chapter 1 Product Overview SM0105 1 6 PMD 2 23 2005 2 54 PM 1 ...
Страница 18: ...2 Graseby MS16A Syringe Driver Chapter 2 Specification and Standards SM0105 2 6 PMD 2 23 2005 2 55 PM 1 ...
Страница 23: ...3 Graseby MS16A Syringe Driver Chapter 3 Circuit Descriptions SM0105 3 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 29: ...4 Graseby MS16A Syringe Driver Chapter 4 Disassembly and Assaembly SM0105 4 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 40: ...5 Graseby MS16A Syringe Driver Chapter 5 Service Test Proceedures SM0105 5 6 PMD 2 23 2005 3 07 PM 1 ...
Страница 50: ...6 Graseby MS16A Syringe Driver Chapter 6 Maintenance SM0105 6 6 PMD 2 23 2005 3 03 PM 1 ...
Страница 55: ...7 Graseby MS16A Syringe Driver Chapter 7 Parts List SM0105 7 6 PMD 2 23 2005 2 59 PM 1 ...