Cautions
Cautions tell you about dangerous conditions that can occur which
may cause damage to the pump if you do not obey all of the
instructions in this manual.
A damaged pump must never be put into active service, as it
may be a hazard to patients through under- or over-infusing.
1.
To prevent serious damage to the driver it must
not be immersed in any liquids or exposed to
strong organic solvents. If immersion occurs or the driver is subject to a serious spillage or
wetting incident, it must be thoroughly tested by a competent person in accordance with the
procedures in this service manual before re-use with a patient.
2.
Wipe off spills immediately. Do not allow fluid or residues to remain on the pump.
3.
The pump is not designed to be autoclaved, steam-sterilised, ETO sterilised or subjected to
temperatures in excess of 45° C (113° F). Failure to observe this caution may cause serious damage
to the pump.
4. To avoid possible malfunction of the pump, do not expose the pump to X- rays, gamma rays or
other ionizing radiation, or to the RF interference or strong electric / magnetic fields emitted (for
example) by arc welding equipment or mobile transmitting equipment.
5.
Moving parts of the driver do not require lubrication during their lifetime. Worn or stiff parts
should be replaced.
6.
Handling the printed circuit board is required during disassembly/ assembly. A static controlled
work station including a conductive mat and grounded wrist strap should be used to provide
protection against electrostatic discharge (ESD) or circuit board damage could result.
Take care to avoid contamination of the exposed switches, circuit tracks and components, e.g. by
the entry of solder particles.
7.
Refer all service, repair, testing and calibrations to suitably qualified technical personnel.
Unauthorised modifications to the driver must not be carried out.
8.
Federal law restricts these Syringe Drivers for use by or on the order of a physician. This applies to
drivers distributed
only within the United States.
SM0105-CON-6.PMD
2/23/2005, 12:12 PM
8
Содержание Graseby MS16A
Страница 13: ...1 Graseby MS16A Syringe Driver Chapter 1 Product Overview SM0105 1 6 PMD 2 23 2005 2 54 PM 1 ...
Страница 18: ...2 Graseby MS16A Syringe Driver Chapter 2 Specification and Standards SM0105 2 6 PMD 2 23 2005 2 55 PM 1 ...
Страница 23: ...3 Graseby MS16A Syringe Driver Chapter 3 Circuit Descriptions SM0105 3 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 29: ...4 Graseby MS16A Syringe Driver Chapter 4 Disassembly and Assaembly SM0105 4 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 40: ...5 Graseby MS16A Syringe Driver Chapter 5 Service Test Proceedures SM0105 5 6 PMD 2 23 2005 3 07 PM 1 ...
Страница 50: ...6 Graseby MS16A Syringe Driver Chapter 6 Maintenance SM0105 6 6 PMD 2 23 2005 3 03 PM 1 ...
Страница 55: ...7 Graseby MS16A Syringe Driver Chapter 7 Parts List SM0105 7 6 PMD 2 23 2005 2 59 PM 1 ...