Chapter 6 - Maintenance
Caution:
Handling of printed circuit boards is required during disassembly/
assembly. A static controlled work station including a conductive mat
and grounded wrist strap should be used to provide protection against
electrostatic discharge (ESD) or circuit board damage could result.
Maintenance
The Department of Health, in its publication MDA DB 9503, has
recommended that infusion drivers are checked by a qualified
technician at regular intervals.
Smiths Medical supports this policy, and recommends that the tests
and if necessary the adjustments given in Chapters 5 and 6 of this
publication should be carried out at regular intervals.
Additionally, with reference to the Department of Health publication
‘SAB(94)26 July 1994’, we recommend that for any of our products
that have been dropped or subjected to severe fluid spills, should be
removed from use immediately, and be checked by a qualified
technician.
Cleaning
Warnings: To prevent serious damage to the driver it must
not be immersed in
any liquids or exposed to strong organic solvents.
Wipe off spills immediately. Do
not allow fluids or residues to remain on the
pump.
The pump is
not designed to be sterilised.
Failure to observe these warnings may cause internal damage to the pump
resulting in patient injury or death.
To clean the driver, wipe the case with a damp cloth (soapy if
necessary). Allow the driver to thoroughly dry before use.
To reduce wear on the actuator assembly, clean the leadscrew using a
small brush with stiff bristles, e.g. a tooth-brush.
Annual performance check
Check the performance of the driver annually following the tests
shown in
Chapter 5, Testing.
SM0105-6-6.PMD
2/23/2005, 3:03 PM
3
Содержание Graseby MS16A
Страница 13: ...1 Graseby MS16A Syringe Driver Chapter 1 Product Overview SM0105 1 6 PMD 2 23 2005 2 54 PM 1 ...
Страница 18: ...2 Graseby MS16A Syringe Driver Chapter 2 Specification and Standards SM0105 2 6 PMD 2 23 2005 2 55 PM 1 ...
Страница 23: ...3 Graseby MS16A Syringe Driver Chapter 3 Circuit Descriptions SM0105 3 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 29: ...4 Graseby MS16A Syringe Driver Chapter 4 Disassembly and Assaembly SM0105 4 6 PMD 2 23 2005 2 56 PM 1 ...
Страница 40: ...5 Graseby MS16A Syringe Driver Chapter 5 Service Test Proceedures SM0105 5 6 PMD 2 23 2005 3 07 PM 1 ...
Страница 50: ...6 Graseby MS16A Syringe Driver Chapter 6 Maintenance SM0105 6 6 PMD 2 23 2005 3 03 PM 1 ...
Страница 55: ...7 Graseby MS16A Syringe Driver Chapter 7 Parts List SM0105 7 6 PMD 2 23 2005 2 59 PM 1 ...