Te c h n i c a l Ser vice Manual
I s s u e 6 ( February 2 0 0 5 )
4 — 1 1
S m i t h s M e d i c a l
4
M S 1 6 A S y r i n g e D r i v e r
f.
Clean off any excess sealant before it sets. The setting time is as
specified for the sealant selected.
Note
: Ensure that the serial number label (item 4) is fitted and
complete.
g.
Fit the battery, observing the correct polarity as shown by the
illustration in the battery compartment.
The alarm sounds for about 15 seconds. If this does not happen,
fit a new battery.
h. Refit the battery compartment cover.
j.
Complete tests specified in Chapter 5,
Testing
.
SM0105-4-6.PMD
2/23/2005, 2:56 PM
11